Water Analysis Instruments

Online Total Arsenic Analyzer

Model: :ADG690

The ULT200 ultrasonic level transmitter provides reliable non-contact level measurement for liquids, solids and slurries. With advanced echo processing technology, it delivers high accuracy even in challenging industrial environments.

  • 24-Hour Response Fast turnaround time
Get a Quote

Model

ADG690

ADG691

ADG692

ADG693

Detection Principle

Arsenomolybdenum Blue Spectrophotometry (Type II)

Cadion Spectrophotometry

DPD Spectrophotometry

4-Aminoantipyrine Extraction - Spectrophotometry

Measuring Range

0-0.2/1/10 mg/L;(Extendable)

0-20 mg/L

0-0.2/1/5/20mg/L

Accuracy

—

±5%or ±0.005mg/L

±5%

—

Repeatability

—

≤5%

—

—

Indication Error

±5%

±10%

-

When ≤0.02 mg/L, ≤0.001 mg/L.At0.02 mg/L, ≤ 5%

≤ ±0.001 mg/L

Lower Limit of

Quantification (LLOQ)

≤0.03mg/L

—

≤0.03 mg/L

0.002mg/L

Zero Drift

±1%

±5%

≤3%

0.001mg/L

Span Drift

±5%

≤5%

±5%

Precision

≤1%

—

≤5%

≤2%

Voltage Stability

±1%

—

±5%

—

Ambient Temperature Stability

±3%

—

Carry-over

±5%

—

Ionic Interference

±5%

—

Data Availability

≥95%

—

Real Sample

Correlation

≤15%

TCd<0.05 mg/L, ±0.005 mg/L

Residual Chlorine≤0.5mg/L

≤ ±0.002 mg/L for concentrations

≤ 0.02 mg/L

 

 

TCd ≥0.05mg/L, ±5%

Residual Chlorine>0.5mg/L

≤ ±10% of reading for

concentrations > 0.02 mg/L

Minimum

Maintenance Interval

≥720h/cycle

—

≥720h/cycle

—

Consistency

≤1%

—

Measurement Time

≤55min

Minimum Measurement Cycle:  40

min/cycle

≤45 min

≤57min

Dimensions (H×W×D)

3 6 0 m m * 2 5 0 m m * 5 0 0 m m ( W * D * H ) + (including pre-treatment unit)

Weight

Net Weight<18Kg;Shipping Weight<20Kg

Environmental

Conditions

Temperature:5 - 40℃:Humidity:≤90%

Power Supply

AC(220±22)V ,(50±1) Hz

Digital

Communication

Serial:   1 × RS232, 1 × RS485 (MODBUS protocol supported) Network:   1 × RJ45 Ethernet

Outputs

Analog: 1 × Isolated 4-20mA Digital: 1 ×Switch/Dry Contact, 1 × DC 24V Sourcing Output

 

 

 

 

 

 

Key Features

 

 

Certified by China National Environmental Monitoring Centre for suitability; ensures high accuracy and data reliability.

Low reagent consumption and waste  generation, reducing operational costs.

Optimized reagent formulation masks interfering factors for more accurate  results.

Supports remote diagnostics and

maintenance; fault reporting via WeChat scan for precise, rapid after-sales service. Compliant with dynamic controltechnical requirements.

Compact design for easytransportation and installation.

 

 

Certified by China National Environmental Monitoring Centre for suitability; ensures monitoring data accuracy and validity.

Low reagent consumption and waste  generation, reducing operational costs.

Optimized reagent formulation masks interfering factors for more accurate  results.

Supports remote diagnostics and

maintenance; fault reporting via WeChat scan for precise, rapid after-sales service. Compliant with dynamic controltechnical requirements.

Compact design for easy transportation and installation.

 

 

Dual-parameter monitoring: smart one- touch switching between Total Chlorine and Residual Chlorine.

Developed based on standard DPD

colorimetric method, eliminating frequent calibration and pH interference issues

common with sensor probes.

Unaffected by temperature, turbidity, or flow rate variations, ensuring high

measurement accuracy.

Extended reagent shelf life of 1 month

(versus typical 2 weeks), reducing

maintenance workload and reagent costs.

High accuracy: Complete pretreatment (distillation + extraction) ensures

measurement sensitivity and precision.

Low limit of quantification (LLOQ), suitable for diverse application scenarios. Safe

operation: Unique, routine low-toxicity extraction and enrichment technology guarantees safety.

Strong anti-interferencecapability: Distillation pretreatment minimizes turbidity and color interference.

Excellent HMI: 10.4-inch color industrial touchscreen.

Low operational costs: Minimal reagent use and waste generation.

Supports remote maintenance: Fault

reporting via WeChat scan for targeted after-sales service.

Compact design for easy transportation and installation.

Typical Applications

Surface water; effluent from industries such

as glass, wood, tanning, textile, chemical,

pottery, pigment, and fertilizer.

Industrial wastewater from metal smelting, metal manufacturing,electroplating,

chemical, and electronics industries.

Mainly used for total (residual) chlorine

monitoring in drinking water, industrial

wastewater, medical wastewater, and

industrial circulating cooling water.

Monitoring stations for surface water

bodies such as rivers, lakes, and reservoirs;

enterprise pollution source discharge

outlets.

Certifications &

Approvals

Software Copyright Registration

Message
Contact number

+86 18961557757

+86 18018362100

Back to top