Model: :ADG18
The ULT200 ultrasonic level transmitter provides reliable non-contact level measurement for liquids, solids and slurries. With advanced echo processing technology, it delivers high accuracy even in challenging industrial environments.
|
Product Model |
ADG18 |
|||
|
Measurement Parameters |
Reagent Model |
ppm CaCO3 |
mmol/l |
|
|
Total Hardness |
Y01 |
0.53 - 5.34 |
0.0053 - 0.0534 |
|
|
Y02 |
1.60 - 16.0 |
0.016 - 0.160 |
||
|
Y03 |
5.34 - 53.4 |
0.0534 - 0.534 |
||
|
Y04 |
16.0 - 160 |
0.160 - 1.60 |
||
|
Y05 |
26.7 - 267 |
0.267 - 2.67 |
||
|
Y06 |
53.4 - 534 |
0.534 - 5.34 |
||
|
Total Alkalinity |
J01 |
4.69 - 187.6 |
0.094 - 3.752 |
|
|
J02 |
7.04 - 281.6 |
0.141 - 5.632 |
||
|
J03 |
9.38 - 375.2 |
0.188 - 7.504 |
||
|
J04 |
14.07 - 562.8 |
0.281 - 11.26 |
||
|
Other Technical Parameters |
Power Supply Voltage |
85-265Vac(47-63Hz) |
||
|
PowerConsumption |
25VA(operation) ,3.5VA(standby) |
|||
|
Protection Rating |
IP54 |
|||
|
Ambient Temperature |
10°C–40°C |
|||
|
Sample Water Temperature |
5°C–40°C |
|||
|
Humidity |
20 -90 % RF (non-condensing) |
|||
|
Sample Water Pressure |
ca. 0.5 - 5 bar (max) (Recommended: 1 - 2 bar) |
|||
|
Inlet Water Requirements |
Clear, colorless, free of solid particles and bubbles |
|||
|
Water Quality Requirements During Hardness Measurement |
pH: 4 – 10; Iron: < 3 ppm; Copper: < 0.2 ppm;Aluminum: < 0.1 ppm; Manganese: < 0.2 ppm;Acid Capacity: KS 4.3 < 5 mmol/l |
|||
|
Installation |
Indoor wall mounting |
|||
|
Dimensions / Weight |
470x370x180 mm (W x H x D)Approx. 6.5 kg |
|||
|
Measurement Method |
Titration colorimetry |
|||
|
Accuracy |
Measurement Accuracy: ±5% of the upper limit for each reagent |
|||
|
±2.5% of the upper limit for each reagent |
||||
|
Reagent Consumption |
Approx. 0.20 ml/cycle (depending on water hardness) |
|||
|
Measurement Time |
Approx. 3 minutes (depending on water hardness) |
|||
|
Usage Count |
Approx. 10,000 analyses for low hardness water. Actual count depends on water hardness and reagent used. |
|||
|
Reagent Shelf Life |
Approx. 2 years |
|||
|
Water Consumption |
Approx. 1 L/cycle at 2 bar pressure. Consumption varies with pressure and rinse time. |
|||